CE Light Industrial

Residential/Light Industrial Immunity and Emissions Requirements per The Generic Standards

 

 

Applicable Standard

Description

Test or Measurement

Required by EN 61000-6-1: 2001

Required Criterion

Electrostatic Discharge

IEC 61000-4-2

 

Air Discharge  ESD at 8 kV  and Contact Discharge at 4 kV at several locations on and around the chassis, with 10 positive and 10 negative ‘hits’ each

Air

Contact

 

8KV

4KV

B

Radiated Immunity

IEC 61000-4-3

 

 

Subject to electromagnetic field of 3V/m from 80 MHz to 1 GHz

Modulated   Frequency

Field

 

 

1 kHz 80% AM

80 – 1000MHz

3 V/m

 

 

A

 

 

 

 

Electrical Fast Transient/Burst

IEC 61000-4-4

 

EFT/Burst of 1.0 kV is coupled to the Power line, any process lines and 0.5 kV on  all I/O lines > 2m.

AC Line (Direct Injection)

Process Lines (Cap. Clamp)

Signal Lines > 3m (Cap. Clamp)

1 kV

 

1 kV

 

0.5 kV

B

Surge Transient

IEC 61000-4-5

 

5 Positive and 5 negative surges are applied to the power terminals  at 0, 90 and 180

Common Mode

Diff. Mode

2 kV

1 kV

B

B

 

Conducted Immunity

IEC 61000-4-6

 

 

3 VRMS is coupled to Power and all I/O & control signals at  0.15 to 80 MHz.

 

 

Power Line / Signal Lines

Frequency

Field

Modulation

0.15 – 80  MHz

3 V

80% AM, 1 kHz

A

 

 

Power Frequency Magnetic Field

IEC 61000-4-8

Magnetically susceptible apparatus is subject  to power freq. H-Field

Hz A/m

50 or 60

30

CRT Interference allowed above 3 A/m

A

Voltage Dips and Interruptions

IEC 61000-4-11

 

Voltage on the Power Lines Is reduced by 30 and 60 %. Voltage on the Power Lines Is interrupted for 5 seconds

Dips

 

 

 

 

Interruptions

30%, 10ms

 

60%, 100ms,

 

 

95%, 5000ms

B

 

C

 

 

C

 

 

Applicable Standard

Test Or Measurement

Required by

EN 6000-6-3: 2001

Notes:

Conducted Emissions

CISPR 22

Phase And Neutral Power Lines

0.15 – 30.0 MHz

Class B Limits

 

Declaration to Class A Permitted for Non-residential

Applications

Radiated Emissions

CISPR 22

Horizontal And Vertical Polarization

30-1000 MHz Class B Limits

Declaration to Class A Permitted for Non-residential

Applications

Power line harmonics

EN61000-3-2

The Power-Line Harmonics

Maximum 1.08, 2.3, 0.43, 1.14, 0.3, 0.77, 0.23 A…. for 2nd to nth Harmonic

 

 Required only for Residential Applications as defined in scope of Applicable Standards

Power line Fluctuations

EN61000-3-3

 

The Power-Line Harmonic Distortion

Maximum Voltage Flicker of   PST < 1, PLT < 0.65

 Required only for Residential Applications as defined in scope of Applicable Standards